Working principle of pressure gauge calibrator

During long-term use, pressure gauges may experience measurement errors due to fatigue of elastic components, wear of transmission mechanisms, and chemical corrosion. Therefore, it is necessary to regularly calibrate the instruments, and new instruments should also be calibrated before installation and use to more appropriately estimate the reliability of the instrument's indicated values. The most commonly used method currently is the comparison method, which uses instruments such as pressure gauge calibrators and precision pressure gauges. So it is crucial to understand the principle of a pressure gauge calibrator and how to use it to calibrate pressure gauges in daily production.

一、Working principle of pressure gauge calibrator

        The pressure gauge calibrator is filled with transformer oil (below 6MPa) or castor oil in a sealed container. Rotate the handwheel to push the piston forward, creating a pressure on the oil that is transmitted in all directions in a closed system. Therefore, the pressure entering the standard instrument, calibrated instrument, and standard instrument is equal. Therefore, using the method of comparison, the absolute error of the calibrated instrument can be obtained.

The accuracy of the pressure gauge calibrator mainly depends on the accuracy of precision pressure gauges, ranging from 0.4 to 0.25 levels. High precision requires the use of precision digital pressure gauges, with accuracy levels of 0.05 and 0.02, which can be used to calibrate standard spring tube pressure gauges, transmitters, etc

Verification principle

         Calibration work is the process of comparing the calibrated instrument with the standard instrument under the same conditions. The selection principle of standard instruments is that when the allowable absolute error of the calibrated instrument is a tolerance, the allowable absolute error of the standard instrument should not exceed a tolerance of 1/3 (preferably not exceeding a tolerance of 1/5). In this way, it can be considered that the reading of the standard instrument is the true value. In addition, to prevent overtravel damage to the standard instrument, the measurement range of the standard instrument should be one level larger than that of the calibrated instrument. If the accuracy level of the calibrated instrument is higher than the indicated level, the instrument is qualified. Otherwise, it should be repaired, replaced, or downgraded for use.

三、Verification content

        Verification is divided into on-site verification and laboratory verification. The verification content includes indication value error, variation, and linear adjustment. The specific steps are: first, evenly determine several calibrated points within the range of the calibrated meter (usually 5-6 points, with certain lower and upper limits of the ship's quantity), and then compare the indicated values of the standard meter point by point from small to large (up stroke) until the maximum value. Advance a little further until the pointer slightly exceeds the maximum value, and then perform a verification from large to small (down stroke). Repeat this process 2-3 times, and finally calculate and determine whether the instrument is qualified based on the definition of various technical indicators.

image.png



Time : Aug 30, 2023
Previous : Already the first
Next : Already the first
News Recommended

Pressure Transmitters for Industrial Pressure Measurement

Xinyi Instrument supplies pressure transmitters for process control, hydraulic systems, petrochemical plants, water treatment, HVAC, power generation and general industrial pressure monitoring. Our pressure transmitter range covers gauge pressure, absolute pressure, differential pressure, high temperature media and digital communication applications.

Choose from compact pressure transmitters, smart 3051 differential pressure transmitters, diaphragm seal models, RS485 digital pressure transmitters and high frequency dynamic pressure sensors. Standard outputs include 4-20 mA, voltage output, HART and RS485 Modbus options, with stainless steel wetted parts and custom process connections available on request.

Pressure Transmitter Parameters

Pressure Types Gauge, absolute, negative pressure, differential pressure
Measuring Range From low differential pressure to high pressure ranges up to 100 MPa, depending on model
Output Signals 4-20 mA, 0-5 V, 1-5 V, 0-10 V, RS485 Modbus, HART options
Accuracy Typical options include 0.1%, 0.2%, 0.25% and 0.5% FS
Process Connection M20 x 1.5, G1/4, G1/2, NPT and customized thread connections
Wetted Materials Stainless steel, 316L diaphragm and corrosion-resistant sealing options
Media Water, oil, gas, air, steam and compatible liquid or gas media
Applications Pipeline pressure, tank level, flow differential pressure, hydraulic pressure and automation systems

Recommended Pressure Transmitter Series

  • Xinyi YW-150 - differential pressure transmitter for stable industrial measurement.
  • Xinyi YW-3051DP - smart differential pressure transmitter for process control.
  • Xinyi YW-130 / YW-140 - compact pressure transmitters for general pressure monitoring.
  • RS485 Digital Pressure Transmitter - Modbus communication for remote monitoring systems.
  • High Temperature Pressure Transmitter - designed for hot media and demanding process conditions.
  • Siemens, Yokogawa and ABB Options - international brand transmitters available for project matching.

Pressure Transmitter FAQ

What is a pressure transmitter used for?

A pressure transmitter converts the pressure of liquid, gas or steam into a standard electrical signal for PLC, DCS, recorder or control instrument input. It is widely used for pipeline pressure, tank level, flow measurement and process safety monitoring.

How do I choose the right pressure transmitter?

Confirm the pressure range, pressure type, medium, temperature, output signal, accuracy, installation thread, electrical connection and environmental requirements. For corrosive media, high temperature or sanitary applications, diaphragm material and sealing structure are especially important.

What is the difference between gauge, absolute and differential pressure transmitters?

Gauge pressure transmitters measure pressure relative to atmospheric pressure. Absolute pressure transmitters measure pressure relative to vacuum. Differential pressure transmitters measure the pressure difference between two points and are commonly used for flow, filter and level measurement.

Can Xinyi Instrument provide customized pressure transmitters?

Yes. Xinyi Instrument can support customized pressure ranges, process connections, output signals, cable length, display options and model selection for different industrial applications.